HP (Hewlett-Packard) 8559A Work Light User Manual


 
MODEL 8559A PERFORMANCE TESTS
PERFORMANCE TESTS
4
-
13. RESIDUAL FM (Cont'd)
Comb Generator:
...............................................
COMB FREQUENCY
-
MHz
100 MC
......................................
INTERPOLATION AMPLITUDE
-
1 MHz
OFF
..............................................
OUTPUT AMPLITUDE Full clockwise
2.
Connect OUTPUT of comb generator to spectrum analyzer INPUT 50Q as shown in Figure 4-5.
NOTE
The 8559A is sensitive to vibration. Be sure spectrum analyzer is in
a
vibration
-
free environment.
3. Adjust spectrum analyzer TUNING control to display 3.0 GHz signal produced by comb generator.
Adjust REFERENCE LEVEL and REF LEVEL FINE controls to position peak of signal at top graticule
line.
4.
Keep 3.0 GHz signal centered on CRT with TUNING control while reducing FREQ SPAN/DIV to zero.
5.
Set RESOLUTION BW to 10 kHz and SWEEP TIME/DIV to 0.1 sec.
6.
Slightly readjust spectrum analyzer FINE TUNING control until trace appears between fourth and
seventh graticule lines. Peak
-
to
-
peak variation of trace should not exceed
1.4
vertical division for each
horizontal division (see Figure
4
-
7).
div
NOTE
For 2201240 line voltages, peak
-
to
-
peak variation of trace should not exceed
1.4 vertical divisions (2
kHz)
in a 180
-
series display mainframe.
FREQUENCY VARIATION
OF IF SIGNAL
AMPLITUDE
VARIATION OF
IF SIGNAL
FIGURE
44.
RESIDUAL FM TO AM CONVERSION DISPLAY FIGURE
4
-
7.
RESIDUAL FM DISPLAY