MODEL 8559A PERFORMANCE TESTS
PERFORMANCE TESTS
4
-
23.
REFERENCE LEVEL ACCURACY
SPECIFICATION:
Step Accuracy (steps referenced with
0 dB input attenuation):
-
10 dBm to
-
80 dBm:
+
0.5 dB
-10dBmto-100dBm: +1.0dB
Vernier Accuracy:
f
0.5 dB
DESCRIPTION:
The reference level accuracy is tested over the range of
-
10 dBm to
-
100 dBm by checking the IF gain steps in
1
dB/DIV (Log) and in LIN. The resulting maximum deviation in each case must be less than k0.5 dB from
-
10 dBm to
-
80 dBm and less than
k
1.0 dB from
-
10 dBm to
-
100 dBm.
SPECTRUM
ANALYZER
SIQNAL QENERATOR
ADAPTER
t
STEP
ATTENUATOR
3550
OPT
H82
A
!
!
STEP
ATTENUATOR
S55C
OPT
H8O
FIGURE
4
-
18.
REFERENCE LEVEL ACCURACY TEST SETUP
EQUIPMENT:
SignalGenerator ......................................................... HP8640B
10
-
dB Attenuator ................................................ HP 8491B, Opt. 010
Step Attenuator (1
dB/step) ........................................
HP 355C, Opt. H80
Step Attenuator (10 dB/step) ...................................... HP 355D, Opt. H82
Adapter, Type N (m) to BNC
(f)
(2 required)
...............................
HP 1250
-
0780