HP (Hewlett-Packard) 8559A Work Light User Manual


 
MODEL 8559A
PERFORMANCE TESTS
PERFORMANCE TESTS
4
-
19.
FREQUENCY RESPONSE (Cont'd)
NOTE
The HP 853A Spectrum Analyzer Display is not recommended for use in this
procedure.
The HP
8350A Sweep Oscillator may be substituted for the HP 8620C in this
procedure,
if
necessary.
EQUIPMENT:
Variable
Persistence/Storage Display
.
.
.........................................................
Sweep Oscillator HP 8620C
............................................................
RF Plug
-
in HP 86222A
........................................................
RF Plug
-
in HP 86290B-H08
...........................................................
Power Meter HP 435A/B
...........................................................
Power Sensor HP 8485A
.................................................
Power Splitter
HP 1
1667A, Opt. C16
................................................
Attenuator, 20
-
dB
HP
8491B, Opt. 020
CrystalDetector
........................................................
HP33330C
...............................
Adapter, Type N (m) to SMA
(f)
(2 required)
HP 1250
-
1250
..........................................
Adapter, Type N
(f)
to SMA(f)
HP 1250
-
1745
...........................................
Adapter, Type N (m) to N (m)
HP 1250
-
0778
......................................
Test Cable, SMC (m) to BNC (m)
HP 1 1592
-
60001
...................................
Cable Assembly, SMA (m) to SMA (m)
HP 8120
-
1 578
PROCEDURE:
1.
Set equipment controls as follows:
Spectrum Analyzer:
FREQUENCYBANDGHz
..................................................
.01-3
TUNING
..............................................................
0.000GHz
FREQSPAWDIV
.........................................................
10MHz
RESOLUTION BW
................................................
300 kHz, coupled
INPUTATTEN
..............................................................
OdB
REFERENCELEVEL
....................................................
-20dBm
............................................................
REFLEVELFINE
-
4
Amplitudescale
........................................................
10dB/DIV
SWEEPTIME/DIV
........................................................
AUTO
SWEEPTRIGGER
.....................................................
FREERUN
ALTIF
.....................................................................
OFF
SIGIDENT
.................................................................
OFF
BLCLIP
....................................................................
OFF
VIDEO FILTER
..............................................
Full CW (not in detent)