HP (Hewlett-Packard) 8559A Work Light User Manual


 
Page
1
-
3:
Table
1-1.
HP
85596
Specifications
(1
of
4)
2236A
&
Below Change
"
Residual FMw specification to read as follows:
less than 1 kHz p
-
p for a time interval less than or equal to
.O1
sec, 100/120 line voltages; less than 2 kHz p
-
p, 220/240
line voltages.
Delete the following under
"
Maximum Input (without damage)
Levelstt
:
Peak Pulse Power
+50
dBm (100W, 10 microsecond pulse width, 0.01% duty cycle)
with input attenuation
>=30 dB.
Change
"Gain Compression
w
specification to read as follows:
Gain compression
is
less than 0.5 dB for a 0 dBm input level
with
0
dB input attenuation.
Under
*'Display Fidelity
w
, change the Linear specification to read
as follows:
<+
-
0.1 division over full 8 division deflection.
Change
"
Humidity Range (Operating)
"
to read as follows:
<95%
R.H.
0
-
degrees
C
to +40-degrees
C.
Change
"EMIft
to read as follows:
Conducted and radiated interference
is
within the requirements
of methods
CEO3 and RE02 of MIL STD 461A,
VDE
0871 and CISPR
Publications
1,
2, and 4.
2320A
&
Below Change
"
Residual FMw specification to read as follows:
less than 1 kHz p
-
p in 0.1 second.