Remote Commands
GB1400 User Manual C-41
clock_amp_up [v] and
clock_amp_dn [v]
Clock Amplitude Up/Down. This command pair is used to increment or decrement the
current clock amplitude by specified amount of [v] volts.
Min. Abbr. clock_amp_u
clock_amp_d
Arguments v: <NR2>, in range 0.05 to 1.50 in steps of 0.05.
Example clock_amp_up 0.05
clock_amp_dn 0.05.
clock_ampl [v]
Clock Amplitude. This command sets the output clock amplitude to [v] volts.
Min. Abbr. clock_ampl
Arguments v <NR2>, in the range 0.50 to 2.00 in steps of 0.05
Example clock_amplitude 2.00
clock_ampl?
Clock Amplitude Query. Returns the current output clock amplitude setting.
Min. Abbr. clock_ampl?
Response: <NR2>
Example: clock_ampl? (command)
CLOCK_AMPL 1.40
clock_off_up and
clock_off_dn
Clock Offset Up, Clock Offset Down (Default). This command pair is used to
increment or decrement the current output clock signal baseline offset by a default value
of 0.05 volts.
Min. Abbr. clock_off_u
clock_off_d
Arguments none