Tektronix 071-0590-00 Welding System User Manual


 
Tutorial
GB1400 User Manual 2-37
D. Locate the controls in the SYNC box of the Analyzer. Verify that the
AUTO SEARCH function is ENABLED. The LED in the AUTO SEARCH key should be ON.
If the LED is OFF, press the AUTO SEARCH function one time to turn the ON. At
this point, verify that the green LOCK LED is ON.
The Analyzer will now search and calculate a new data threshold. What is the
value of the threshold selected by Auto-Search? You should expect to see this
threshold value within a few hundred millivolts of the “data signal mid-point (or
optimum threshold).
Note - Common Setup Problems
Connecting the Generator’s clock output to the Analyzer’s not-clock (clock-
bar) input
Connecting the Generator’s data output to the Analyzer’s not-data (data-bar)
input
Changing Generator’s CLOCK amplitude/offset when lab calls for
adjustments to DATA signals
Failure to connect the external DATA THRESHOLD cable from rear of unit
to NOT_DATA input.
This Concludes the Tutorial.