Remote Commands
GB1400 User Manual C-43
Min. Abbr. data_amp_u
data_amp_d
Arguments none
Example data_amp_dn
data_amp_up [v] and
data_amp_dn [v]
Data Amplitude Up, Data Amplitude Down. This command pair is used to increment
or decrement the current amplitude of the Generator data output by a specified step size
of [v] volts.
Min. Abbr. data_amp_u
data_amp_d
Arguments v: <NR2>, in the range 0.05 to 1.50, in steps of 0.05.
Example data_amp_up 1.00
data_ampl [v]
Data Amplitude. Sets the amplitude of the Generator's data output signal to [v] volts.
Min. Abbr. data_ampl
Arguments v: <NR2>, in the range 0.50 to 2.00 in steps of 0.05.
Example data_ampl 1.00
data_ampl?
Data Amplitude Query. Returns the current amplitude setup of the Generator's data
output signal.
Min. Abbr. data_ampl?
Returns <NR2>
Example data_ampl? (command)
DATA_AMPL 1.00 (response)
data_off_up and
data_off_dn
Data Offset Up/Down (Default). This command pair is used to increment or decrement
the current baseline offset of the Generator data output by a default amount of 0.05 volts.