Tektronix 071-0590-00 Welding System User Manual


 
Remote Commands
C-64 GB1400 User Manual
rdata_thr_up and
rdata_thr_dn
Reference Data Threshold Up/Down (Default). This command pair is used to
increment (rdata_thr_up) or decrement (rdata_thr_dn) the current REF DATA threshold
by the default increment of 0.05 volts.
Min. Abbr. rdata_thr_u
rdata_thr_d
Argument none
Examples rdata_thr_up
rdata_thr_dn
rdata_thr_up [v] and
rdata_thr_dn [v]
Reference Data Threshold Up/Down. This command pair is used to increment
(rdata_thr_up) or decrement (rdata_thr_dn) the current REF DATA threshold by v,
where v is in <NR2> format and may be set in the range 0.05 to 2.50 Volts in 0.05 V
steps.
Min. Abbr. rdata_thr_u
rdata_thr_d
Argument v: <NR2>, 0.05 to 2.50 volts, in 0.05 V steps.
Examples rdata_thr_up 0.10
rdata_thr_dn 0.50
rdata_thres [v]
Reference Data Threshold. Sets the threshold voltage for the Analyzer REF DATA
input to v, where the allowed range for v depends on the current input clock termination.
Min. Abbr. rdata_thre
Arguments v: -1.50 to 1.00 in 0.05 v steps, for terminations
GND, AC
-2.50 to 0.00 in 0.05 v steps, for termination -2V
Example rdata_thres 1.00