Remote Commands
C-42 GB1400 User Manual
Examples clock_off_up
clock_off_dn
clock_off_up [v] and
clock_off_dn [v]
Clock Offset Up, Clock Offset Down. This command pair is used to increment or
decrement the current output clock signal baseline offset by a specified amount [v] volts.
Min. Abbr. clock_off_u
clock_off_d
Arguments v: <NR2>, in the range 0.05 to 3.00 in 0.05 v steps.
in the range 0.05 to 3.80 in 0.05 v steps (PECL option)
Example clock_off_up 0.10
clock_off_dn 0.10
clock_offset [v]
Clock Offset. Sets the baseline offset of the output clock signal to [v] volts.
Min. Abbr. clock_offs
Arguments v: <NR2>, in the range -2.00 to 1.00 in steps of 0.05.
in the range -2.00 to 1.80 in 0.05 v steps (PECL option)
Example clock_offset -0.75
clock_offset?
Clock Offset Query. Returns the current value of the output clock signal baseline
offset.
Min. Abbr. clock_offs?
Returns <NR2>
Example clock_offset? (command)
CLOCK_OFFSET -0.75 (response)
data_amp_up and
data_amp_dn
Data Amplitude Up/Down (Default). This command pair is used to increment or
decrement the current amplitude of the Generator data output by a default amount of 0.05
volts.