Tektronix 071-0590-00 Welding System User Manual


 
Remote Commands
GB1400 User Manual C-61
data_thr_up and
data_thr_dn
Data Threshold Up/Down (Default). This pair of commands increments (data_thr_up)
or decrements (data_thr_dn) the current Analyzer DATA input threshold by the default
increment of 0.05 volts.
Min. Abbr. data_thr_u
data_thr_d
Arguments none
Examples data_thr_up (increments threshold by 0.05 volts)
data_thr_dn (decrements threshold by 0.05 volts)
data_thr_up [v] and
data_thr_dn [v]
Data Threshold Up/Down. This pair of commands increments (data_thr_up) or
decrements (data_thr_dn) the current Analyzer DATA input threshold by an amount v.
Min. Abbr. data_thr_u
data_thr_d
Arguments v: 0.05 to 2.50 in 0.05 V steps
Examples data_thr_up 0.10 (increments threshold by 0.1 volt)
data_thr_dn 2.00 (decrements threshold by 2 volts)
data_thres [v]
Data Threshold. Sets the threshold voltage for the Analyzer DATA input to v, where
the allowed range for v depends on the current DATA input termination.
Min. Abbr. data_thre
Arguments v: -1.50 to 1.00 in 0.05 v steps, for input terminations
GND, AC
-2.50 to 0.00 in 0.05 v steps, for input terminations -2V
Example data_thres -1.50